• $3,147.99
  • Free 30-day Returns
  • Price alert
    Shop it at NeweggBusiness.

    Meet Your Seller

    Fluke 725 Multifunction Process Calibrator

    • Brand; Fluke
    • Fluke 725
    • Batteries : 4 AA batteries required. (included)
    • model number : FLUKE-725 US
    • Weight: 1.22 Kilograms
    • Dimensions LxWxH; 10 x 7.8 x 3.3 inches
    +
    +
    Overview
    Specs
    Reviews
    avatar

    Any questions? Our AI beta will help you find out quickly.

    The Fluke 725 Multifunction Process Calibrator is a powerful, yet easy-to-use field calibrator. Use the measure and source functions to test and calibrate almost any process parameter
    Small, streamlined shape, rugged, reliable design stands up to field use
    Easy to read measure/source screen lets you view input and output simultaneously
    Measure volts, mA, RTDs, thermocouples, frequency, and ohms to test sensors and transmitters
    Source/simulate volts, mA, thermocouples, RTDs, frequency, ohms, and pressure to calibrate transmitters

    The multifunction process calibrator is a powerful yet easy-to-use field calibrator. Use the measure and source functions to test and calibrate almost any process parameter. Source/simulate volts, mA, thermocouples, RTDs, frequency, ohms, and pressure to calibrate transmitters. Support flow meter testing with frequency and CPM functions. Perform fast linearity tests with auto- step and auto-ramp features. Power transmitters during test using loop supply with simultaneous mA measurement. Handling of fast pulsed RTD transmitters and PLCs, with pulses as short as 10 ms. Store frequently used test setups for later use.

    Backlight lets you work in poor light. Large battery capacity of four AA cells. Battery door for easy changes.
    Voltage range: 0 to 100 mV, 0 to 10V (source), 0 to 30V (measure); voltage resolution - 0.01 mV, 0.01V, 0.01V; voltage accuracy - 0.02 percent Rdg + 2 LSD. Notes Max load, 1
    mA. mA: 0 to 24 range, 0.001 mA resolution,. 02 percent Rdg + 2 LSD accuracy. Notes max load, 1000 ohm. mV (TC terminals): - 10.00 mV to 75.00 mV range,. 01 mV resolution,. 025 percent of range + 1 LSD accuracy.
    Resistance: 15 to 3200 ohm range, 0.01 to 0.1 ohm resolution, 0.10 to 1.0 ohm.
    Frequency range: 2.0 to 1000.0 CPM, 1 to 1000 Hz, 1.0 to 10.0 kHz; Frequency resolution - 0.1 CPM, 1 Hz, 0.1 kHz; Frequency accuracy - +/-. 05 percent, +/-. 05 percent, +/-. 25 percent. For frequency source, waveform is 5V p-p squarewave, - 0.1V offset.
    Loop Supply: 24V dc range, 10 percent accuracy. 30V maximum voltage. -40 to 71 degree C storage temperature. -10 to 55 degree C operating temperature.
    Relative humidity: 95 percent (10 to 35 degree C), 75 percent (30 to 40 degree C), 45 percent (40 to 50 degree C), 35 percent (50 to 55 degree C).
    Shock: 30g, 11ms, half-sine shock (or 1meter drop test).
    Vibration: random, 2g, 5-500
    Hz. safety: CSA C 22.2
    No. 1010.1: 1992.
    Battery: Four AA alkaline batteries.
    Battery life: 25 hours typical.

    Warranty & Returns

    Warranty, Returns, And Additional Information

    Warranty